Measurement Parameters |
Resistance Unbalance Capacitance Unbalance - |
DR1, DR2, DR3 K1, K2, K3 E1, E2, E3 |
Measuring Ranges |
Resistance Unbalance Capacitance Unbalance - |
-19.99% to +19.99% -1999 to +1999 pF & -19.99 to +19.99 nF in two ranges |
Indication |
3 ½ digit panel meter with sign |
|
Resolution |
Resistance Unbalance Capacitance Unbalance |
0.01% 1pF |
Test Voltage |
Resistance Unbalance Capacitance Unbalance - |
2V DC 6V AC with suitable frequencies |
Accuracy |
Resistance Unbalance Capacitance Unbalance |
+ 0.01 % + 1% + 1pF + 1digit |
Minimum Loop Resistance |
1 Ohm. |
|
Selection of Parameters |
By means of rotary switch on front panel |
|
Power Supply |
230V ± 15%, 60Hz. |
|
Dimensions |
435(L) x 135(H) x 305(W) mm |
|
Weight |
8 Kg. Approx. |