This set includes Bridge Model:MLS11DS & HV Power Supply model :HLS12SA necessary cables & very flexible, user friendly.Temp correction for the tan delta of object is automatically done as per IEC.Suitable for measuring C & Tan delta for generators, transformers, bushings, circuit breakers,cables, motors at various test voltages upto 12kV.
PARAMETER | RANGE | RESOLUTION | ACCURACY |
---|---|---|---|
CAPACITANCE | 1pF to 1.1mfd | 0.01pF | ±0.2% of the reading ±1 pF in UST Mode |
TAN DELTA | 0 % – 100% | 0.001%(0.00001) | ± 1% of the reading ± 0.0002 |
VOLTAGE Measurement | 100V –12 KV | 1.0V | ±1% of reading ±1digit |
CURRENT Measurement | 0 –200 mA | 1mA | ±1 %, ±1digit |
POWER FACTOR | 0-1 | 0.01%(0.0001) | ±0.1% of the reading ±0.0002 |
POWER INPUT | 230 Volts±10%,50Hz . Single phase. |
POWER OUTPUT | 0-12KV , 200mA |
Temp. Correction | IEC Temperature Correction for Tan Delta Values. |
DATA STORAGE | More than 100 results storage memory. |
INDICATION | Large 4 line LCD. |
INTERFACE | RS232 PORT, |
KEY BOARD | Multikey, alphanumeric Keypad. |
SOFTWARE | PC side Windows based analysis software for downloaded data from Instrument. |
OTHER PARAMETERS | Apparent Power, True Power, Reactive Power available pc side report. |
IN BUILT STD CAPACITOR SF6 | Capacitance- 100pF ± 0.5%, Tan d< 1 x10-4 |
RESONATING INDUCTOR | Optional useful to increase the current rating suitable to measure 1.1 mFd@ 12KV by expanding current capcity upto 2Amp if requested. |
PROTECTIONS | Zero Start, Interlocks, Open Ground, Over Current, Over Voltage,Over Temperature, Foot switch for HV ON. |
CABLES | HV Cable – 20 M length,12KV ,Double screened with hooks & clamps. LV Cable – 20 M length, screened cable, with hooks & clamps.-2Nos. Grounding Cable – 20M & 10M length with clips –2Nos. Footswitch – Footswitch with 5 Meter length cable. Interconnecting Cables, Mains chord, PC& Printer Cable. |
CALIBRATOR BOX | With one Capacitance & 3 Tan Delta values.Test Voltage 3KV. |
CARRING CASES | Suitable foam padded carring cases for Bridge & Power Supply.Carrying Bag for Cables |
TEMP & HUMIDITY | 10° C – 50° C with HUMIDITY : 30% Rh -90% Rh. |
DIMENSIONS | Power Supply : 540mm X 350mm X310mm Bridge : 540mm X 350mm X220mm |
WEIGHT | Less than 50kg without Leads. |