CAPACITANCE TAN DELTA & DIELECTRIC CONSTANT
Test Voltage
Adjustable From 0 to 2500 VAC in steps with an accuracy of ± 1.0 %± 1 volt
Resolution
1 Volt
Test Frequency
50 Hz
Dielectric Constant Range
1.0 to 20.0 accuracy ± 0.1%, Best Resolution : 0.001
Dielectric Tan Delta Range
0.01% to 100%
Accuracy
±1% ± 10digits, Best Resolution: 0.00001
Dielectric Loss
0 to 10.00 accuracy: ±1% ± 1 digit.
RESISTANCE & RESISTIVITY
Test Voltage DC
100- 1000 VDC in selectable steps in 100V with an accuracy of ± 1.0 %
Resistance Range
10e W-10e W Resolution: 0.01.
Resistance Accuracy
± 2%of reading up to 1GW & ± 5% above 1 GW.
Resistivity Range
10e W-10e W cm Best Resolution: 0.01.
Resistivity Accuracy
± 2% of reading at 10e W-10e W cm
± 5% of reading at 10e W-10e W cm
± 10% of reading above 10e W cm
OIL TEST CELL
Construction
3 Terminal Configuration.
Cell capacitance
55pf± 1 pF.
Material
Stainless steel with Teon spacers.
Electrode Spacing
2mm.
Volume
45ml
OIL TEST CELL HEATER
Power Supply
230VAC ± 10%,50Hz.
45ml
45ml
45ml
45ml
45ml
45ml
Test Temperature
Adjustable from 20°C(Room Temp.) to 15°C with an accuracy ±0.5 % Accuracy of temp. control: ± 2°C.
DISPLAY PRINTER & DATA STORAGE
Display
Alphanumeric LCD display.
Keyboard
20 Key keypad providing numeric and soft function entries.
Printer
Inbuilt dot matrix printer
PC Compatibility
RS 232 port
Data Storage
upto 250 test results in inbuilt memory.
Standard Calibrator
Capacitance 1 value, Tan Delta 3 value & Resistivity 4 value standard calibrator with traceable calibration certicate operating voltage 500VAC/DC.
POWER SUPPLY, TEMP & HUMIDITY RANGE
Power Supply
230VAC ± 10%,50Hz
Operating Temp
10 to 50°C
Humidity
Ambient to 90%RH
Connectors
Suitable HV, LV & Ground Leads ,Safety Interlock & 3Pin Mains chord.