MLS11DA & HLS5A

Technical Features
Test Voltage
0 to 5 kV
Rated Current
100mA Continuous
200mA Short term
Automatic & software controlled voltage setting to avoid abrupt voltage rise.
Automatic selection of seven different modes of operation for measurement of 3 terminal object from windows based software .
1) UST-Ungrounded Specimen Test – 3 Combinations
This mode is used when the object under test is not grounded.
This test provides most accurate results.
1) GST-Grounded Specimen Test
This mode is used when the object under test is permanently grounded. This test is used more often in outdoor installations, power systems etc.
1) GST-Grounded Specimen Test with guard. 3 Combinations
This mode is used for measuring them stray capacitances and separating them from basic measurement in GST mode.
Standard Capacitor
100pF, 6kV built-in
Input Frequency
From 50 Hz ( lines frequency) , ± 5%.
Power Required
230 VAC ± 10%, 1 phase AC , 50Hz.
Safety Features
Zero start Control
Open Ground Indication Lamp with double grounded connection
External Interlock
HT cut-off on overload
HT ON & Supply ON Indication
All operating controls at earth potential
Necessory terminals & Sockets for connection to Bridge
Safety Specifications
IEC 61010 complied.
Vibration Suitability
ASTM D 999 complied.
EMI/EMC Specification
Conducted & Radiated Emission EN 61326 complied,
Radiated Susceptibility
IEC61000-4-3 complied,
Electrical Fast Transients
IEC61000-4-4 complied
Surge Test
IEC61000-4-5 complied,
Conducted Susceptibility
IEC61000-4-6 complied,
Power Freq. Magnetic Field Immunity
ImmunityIEC61000-4-8 complied
Voltage dips / Interuption
IEC61000-4-11 complied.